# Drying & Heating Chambers

- Link: https://flexitest.in/packaging-testing-equipments/drying-heating-chambers/
- Published: 2024-09-14T07:09:12+05:30
- Author: chetanvarne

# Drying & Heating  Chambers

 1. [Home](https://flexitest.in)
 2. [Packaging Testing Equipments](https://flexitest.in/packaging-testing-equipments/)
 3. Drying & Heating Chambers

![Drying & Heating Chambers](https://flexitest.in/wp-content/uploads/2025/04/ED56-
FD56-VD56-M115-main-Banner-scaled.webp "Drying & Heating Chambers 17")

## BINDER Solutions:  Select the right product for your specific requirements

BINDER drying chambers, equipped with natural convection or forced convection, are
well-known for their exceptional quality and dependability. Through significant 
redevelopment efforts, these versatile chambers have been further optimised to meet
the specific demands associated with drying or heating samples. Alongside their 
industry-leading energy efficiency and user-friendly operation, the latest APT.line
™ technology ensures remarkable temperature precision, enhancing the attractiveness
of this new generation of incubators.

## Drying & Heating Chambers Machine Solutions

![Drying and Heating Chambers Model M115](https://flexitest.in/wp-content/uploads/
2025/04/M115-e1745826967255.webp "Drying & Heating Chambers 18")

## 󠀁[Drying and Heating Chambers Model M115](https://flexitest.in/packaging-testing-equipments/drying-and-heating-chambers-model-m115/)󠁿

![Vacuum Drying Chambers Model VD 56](https://flexitest.in/wp-content/uploads/2025/
04/VD-56-e1745826845700.webp "Drying & Heating Chambers 19")

## 󠀁[Vacuum Drying Chambers Model VD 56](https://flexitest.in/packaging-testing-equipments/drying-heating-chambers/drying-and-heating-chamber-model-vd-56/)󠁿

![Drying and Heating Chamber Model FD 56](https://flexitest.in/wp-content/uploads/
2025/04/ED56-2-e1745826816151.webp "Drying & Heating Chambers 20")

## 󠀁[Drying and Heating Chamber Model FD 56](https://flexitest.in/packaging-testing-equipments/drying-heating-chambers/drying-and-heating-chamber-model-fd-56/)󠁿

![Drying and Heating Chambers Model M115](https://flexitest.in/wp-content/uploads/
2025/04/ED56-1.webp "Drying & Heating Chambers 21")

## 󠀁[Drying and Heating Chamber Model ED 56](https://flexitest.in/packaging-testing-equipments/drying-heating-chambers/drying-and-heating-chamber-model-ed-56/)󠁿

## Select the right product for your  specific requirements

Technical specifications for BINDER drying and heating chambers — Models M115, VD
56, FD 56, and ED 56.

 ← Swipe to see full table →

 BINDER Drying Chamber Specifications  |  Specification |  Unit |  Model M115 |  Model VD 56 |  Model FD 56 |  Model ED 56 |  
    |  General |  
 |  Type |   |  Forced convection
with prog. functions |  Vacuum dryingchamber |  Forced convectiondrying oven |  Natural convectiondrying oven |  
 |  Interior volume |  L |  113 |  56 |  57 |  57 |  
 |  Temperature range |  °C |  Amb +23 to 300 |  Amb +9 to 220 |  Amb +10 to 300 |  Amb +5 to 300 |  
 |  Convection type |   |  Forced |  Vacuum |  Forced |  Natural |  
  |  Dimensions |  
 |  Interior (W × D × H) |  mm |  550 × 375 × 550 |  400 × 343 × 400 |  400 × 345 × 440 |  360 × 380 × 420 |  
 |  Exterior (W × D × H) |  mm |  710 × 636 × 732 |  638 × 461 × 815 |  560 × 565 × 625 |  560 × 565 × 625 |  
  |  Weight & Load |  
 |  Net weight (empty) |  kg |  58 |  96 |  41 |  42 |  
 |  Permitted load (total) |  kg |  150 |  60 |  70 |  70 |  
 |  Load per rack |  kg |  30 |  20 |  30 |  30 |  
 |  Shelves (std / max) |   |  2 / 5 |  2 / 3 |  2 / 4 |  2 / 4 |  
  |  Electrical |  
 |  Nominal power |  kW |  2.4 |  1.4 |  1.2 |  1.05 |  
 |  Voltage |  V |  230 |  200–230 |  230 |  230 |  
 |  Frequency |  Hz |  50 / 60 |  50 / 60 |  50 / 60 |  50 / 60 |

* Specifications shown are nominal. Final configuration may vary by region and options
selected.

General

Type

Forced convection with prog. functions

Interior volume

113 L

Temperature range

Amb +23 to 300 °C

Convection type

Forced

Dimensions

Interior (W×D×H)

550 × 375 × 550 mm

Exterior (W×D×H)

710 × 636 × 732 mm

Weight & Load

Net weight (empty)

58 kg

Permitted load

150 kg

Load per rack

30 kg

Shelves (std / max)

2 / 5

Electrical

Nominal power

2.4 kW

Voltage

230 V

Frequency

50 / 60 Hz

General

Type

Vacuum drying chamber

Interior volume

56 L

Temperature range

Amb +9 to 220 °C

Convection type

Vacuum

Dimensions

Interior (W×D×H)

400 × 343 × 400 mm

Exterior (W×D×H)

638 × 461 × 815 mm

Weight & Load

Net weight (empty)

96 kg

Permitted load

60 kg

Load per rack

20 kg

Shelves (std / max)

2 / 3

Electrical

Nominal power

1.4 kW

Voltage

200–230 V

Frequency

50 / 60 Hz

General

Type

Forced convection drying oven

Interior volume

57 L

Temperature range

Amb +10 to 300 °C

Convection type

Forced

Dimensions

Interior (W×D×H)

400 × 345 × 440 mm

Exterior (W×D×H)

560 × 565 × 625 mm

Weight & Load

Net weight (empty)

41 kg

Permitted load

70 kg

Load per rack

30 kg

Shelves (std / max)

2 / 4

Electrical

Nominal power

1.2 kW

Voltage

230 V

Frequency

50 / 60 Hz

General

Type

Natural convection drying oven

Interior volume

57 L

Temperature range

Amb +5 to 300 °C

Convection type

Natural

Dimensions

Interior (W×D×H)

360 × 380 × 420 mm

Exterior (W×D×H)

560 × 565 × 625 mm

Weight & Load

Net weight (empty)

42 kg

Permitted load

70 kg

Load per rack

30 kg

Shelves (std / max)

2 / 4

Electrical

Nominal power

1.05 kW

Voltage

230 V

Frequency

50 / 60 Hz

* Nominal specifications. Final configuration may vary.

## FAQ's

Frequently Asked Questions –
Drying & Heating Chambers

  What is a drying and heating chamber used for in packaging testing?

A drying and heating chamber provides precise, uniform temperature control for conditioning,
drying, ageing, and heat-resistance testing of packaging materials. It prepares 
samples before tensile, seal strength, and barrier property testing, in compliance
with ASTM D618 for conditioning plastics and ISO 2233 for conditioning flexible 
packaging materials, paper, and board before mechanical testing.

  What is the difference between natural convection and forced convection drying
ovens?

Natural convection ovens like BINDER M115 use passive airflow, ideal for lightweight
or dust-sensitive packaging samples. Forced convection ovens like BINDER FD 56 and
ED 56 use a motorised fan for faster, more uniform heating. ASTM E145 classifies
both types and defines performance and uniformity specifications for gravity-convection
and forced-ventilation laboratory ovens.

  When should a vacuum drying oven be used instead of a standard heating chamber?

A vacuum drying oven like BINDER VD 56 dries samples at low pressure and lower temperatures,
preventing oxidation and thermal degradation. It is recommended for heat-sensitive
packaging films, hygroscopic pharmaceutical powders, and moisture-sensitive adhesives
where standard oven temperatures per ASTM D618 or ISO 291 would alter the material
structure or chemical composition of the sample.

  Which ASTM standards apply to drying and heating chambers used in packaging testing?

Key ASTM standards include ASTM E145 for gravity-convection and forced-ventilation
oven specifications, ASTM D618 for conditioning plastics before testing, ASTM D4332
for conditioning packaging containers and flexible pouches before testing, and ASTM
F1980 for accelerated ageing of sterile barrier and packaging systems to determine
shelf life and expiry date validation.

  Which ISO standards govern heating chambers used for packaging material conditioning?

Relevant ISO standards include ISO 291 for standard atmospheres for conditioning
and testing plastics, ISO 2233 for conditioning flexible packaging materials, paper,
and board before testing, ISO 188 for accelerated ageing and heat resistance testing
of rubber-based packaging materials, and IEC 60068-2-2 for dry heat environmental
simulation testing of packaging components and materials.

  What is BINDER APT.line™ technology and how does it benefit packaging laboratories?

BINDER APT.line™ is a preheating chamber system that warms incoming fresh air before
it enters the main working space. This prevents cold air intrusion during door openings
and maintains exceptional temperature uniformity — essential for reproducible conditioning
results per ASTM D618 and ISO 2233 when testing packaging films, laminates, adhesive
bonds, and heat-sensitive flexible packaging materials.

  How must packaging samples be conditioned before mechanical testing in a heating
chamber?

Per ASTM D4332 and ISO 2233, packaging samples must be conditioned at a defined 
temperature and relative humidity for a specified minimum period before any mechanical
testing. Forced-convection ovens ensure uniform heat distribution across all samples,
eliminating moisture variability and ensuring that tensile strength, elongation,
and seal integrity test results are accurate and reproducible.

  Can drying ovens be used for accelerated ageing of packaging materials to determine
shelf life?

Yes. BINDER FD 56 and ED 56 forced-convection ovens are widely used for accelerated
thermal ageing of flexible films, laminates, adhesives, and sterile barrier packaging
systems. Testing follows ASTM F1980, which uses elevated temperature ageing to predict
long-term packaging performance and establish expiry dates for pharmaceutical and
medical device packaging without waiting for full real-time ageing data.

  What temperature range and uniformity do BINDER heating chambers provide for packaging
testing?

BINDER heating chambers offer working temperature ranges from ambient up to 300°
C depending on the model. Forced-convection models FD 56 and ED 56 deliver temperature
uniformity within ±0.5°C to ±1°C throughout the working space. This precision meets
ASTM E145 oven uniformity requirements and is essential for reproducible accelerated
ageing per ASTM F1980 and conditioning per ASTM D618.

  What safety and calibration standards apply to BINDER laboratory drying and heating
chambers?

DIN 12880 governs safety requirements for electrically heated laboratory drying 
ovens, covering overtemperature protection, construction, and electrical safety.
For calibration, chambers must meet ISO 17025 traceability requirements. All BINDER
ovens include independent overtemperature protection and are designed in full compliance
with DIN 12880, ensuring safe and reliable operation in packaging, pharmaceutical,
and industrial testing laboratories.