Drying & Heating Chambers

  1. Home
  2. Packaging Testing Equipments
  3. Drying & Heating Chambers
Drying & Heating Chambers

BINDER Solutions: Select the right product for your specific requirements

BINDER drying chambers, equipped with natural convection or forced convection, are well-known for their exceptional quality and dependability. Through significant redevelopment efforts, these versatile chambers have been further optimised to meet the specific demands associated with drying or heating samples. Alongside their industry-leading energy efficiency and user-friendly operation, the latest APT.line™ technology ensures remarkable temperature precision, enhancing the attractiveness of this new generation of incubators.

Select the right product for your specific requirements

Technical specifications for BINDER drying and heating chambers — Models M115, VD 56, FD 56, and ED 56.

← Swipe to see full table →
BINDER Drying Chamber Specifications
Specification Unit Model M115 Model VD 56 Model FD 56 Model ED 56
General
Type Forced convection
with prog. functions
Vacuum drying
chamber
Forced convection
drying oven
Natural convection
drying oven
Interior volume L 113 56 57 57
Temperature range °C Amb +23 to 300 Amb +9 to 220 Amb +10 to 300 Amb +5 to 300
Convection type Forced Vacuum Forced Natural
Dimensions
Interior (W × D × H) mm 550 × 375 × 550 400 × 343 × 400 400 × 345 × 440 360 × 380 × 420
Exterior (W × D × H) mm 710 × 636 × 732 638 × 461 × 815 560 × 565 × 625 560 × 565 × 625
Weight & Load
Net weight (empty) kg 58 96 41 42
Permitted load (total) kg 150 60 70 70
Load per rack kg 30 20 30 30
Shelves (std / max) 2 / 5 2 / 3 2 / 4 2 / 4
Electrical
Nominal power kW 2.4 1.4 1.2 1.05
Voltage V 230 200–230 230 230
Frequency Hz 50 / 60 50 / 60 50 / 60 50 / 60

* Specifications shown are nominal. Final configuration may vary by region and options selected.

General
Type
Forced convection with prog. functions
Interior volume
113 L
Temperature range
Amb +23 to 300 °C
Convection type
Forced
Dimensions
Interior (W×D×H)
550 × 375 × 550 mm
Exterior (W×D×H)
710 × 636 × 732 mm
Weight & Load
Net weight (empty)
58 kg
Permitted load
150 kg
Load per rack
30 kg
Shelves (std / max)
2 / 5
Electrical
Nominal power
2.4 kW
Voltage
230 V
Frequency
50 / 60 Hz
General
Type
Vacuum drying chamber
Interior volume
56 L
Temperature range
Amb +9 to 220 °C
Convection type
Vacuum
Dimensions
Interior (W×D×H)
400 × 343 × 400 mm
Exterior (W×D×H)
638 × 461 × 815 mm
Weight & Load
Net weight (empty)
96 kg
Permitted load
60 kg
Load per rack
20 kg
Shelves (std / max)
2 / 3
Electrical
Nominal power
1.4 kW
Voltage
200–230 V
Frequency
50 / 60 Hz
General
Type
Forced convection drying oven
Interior volume
57 L
Temperature range
Amb +10 to 300 °C
Convection type
Forced
Dimensions
Interior (W×D×H)
400 × 345 × 440 mm
Exterior (W×D×H)
560 × 565 × 625 mm
Weight & Load
Net weight (empty)
41 kg
Permitted load
70 kg
Load per rack
30 kg
Shelves (std / max)
2 / 4
Electrical
Nominal power
1.2 kW
Voltage
230 V
Frequency
50 / 60 Hz
General
Type
Natural convection drying oven
Interior volume
57 L
Temperature range
Amb +5 to 300 °C
Convection type
Natural
Dimensions
Interior (W×D×H)
360 × 380 × 420 mm
Exterior (W×D×H)
560 × 565 × 625 mm
Weight & Load
Net weight (empty)
42 kg
Permitted load
70 kg
Load per rack
30 kg
Shelves (std / max)
2 / 4
Electrical
Nominal power
1.05 kW
Voltage
230 V
Frequency
50 / 60 Hz

* Nominal specifications. Final configuration may vary.

FAQ's

Frequently Asked Questions –
Drying & Heating Chambers

A drying and heating chamber provides precise, uniform temperature control for conditioning, drying, ageing, and heat-resistance testing of packaging materials. It prepares samples before tensile, seal strength, and barrier property testing, in compliance with ASTM D618 for conditioning plastics and ISO 2233 for conditioning flexible packaging materials, paper, and board before mechanical testing.

Natural convection ovens like BINDER M115 use passive airflow, ideal for lightweight or dust-sensitive packaging samples. Forced convection ovens like BINDER FD 56 and ED 56 use a motorised fan for faster, more uniform heating. ASTM E145 classifies both types and defines performance and uniformity specifications for gravity-convection and forced-ventilation laboratory ovens.

A vacuum drying oven like BINDER VD 56 dries samples at low pressure and lower temperatures, preventing oxidation and thermal degradation. It is recommended for heat-sensitive packaging films, hygroscopic pharmaceutical powders, and moisture-sensitive adhesives where standard oven temperatures per ASTM D618 or ISO 291 would alter the material structure or chemical composition of the sample.

Key ASTM standards include ASTM E145 for gravity-convection and forced-ventilation oven specifications, ASTM D618 for conditioning plastics before testing, ASTM D4332 for conditioning packaging containers and flexible pouches before testing, and ASTM F1980 for accelerated ageing of sterile barrier and packaging systems to determine shelf life and expiry date validation.

Relevant ISO standards include ISO 291 for standard atmospheres for conditioning and testing plastics, ISO 2233 for conditioning flexible packaging materials, paper, and board before testing, ISO 188 for accelerated ageing and heat resistance testing of rubber-based packaging materials, and IEC 60068-2-2 for dry heat environmental simulation testing of packaging components and materials.

BINDER APT.line™ is a preheating chamber system that warms incoming fresh air before it enters the main working space. This prevents cold air intrusion during door openings and maintains exceptional temperature uniformity — essential for reproducible conditioning results per ASTM D618 and ISO 2233 when testing packaging films, laminates, adhesive bonds, and heat-sensitive flexible packaging materials.

Per ASTM D4332 and ISO 2233, packaging samples must be conditioned at a defined temperature and relative humidity for a specified minimum period before any mechanical testing. Forced-convection ovens ensure uniform heat distribution across all samples, eliminating moisture variability and ensuring that tensile strength, elongation, and seal integrity test results are accurate and reproducible.

Yes. BINDER FD 56 and ED 56 forced-convection ovens are widely used for accelerated thermal ageing of flexible films, laminates, adhesives, and sterile barrier packaging systems. Testing follows ASTM F1980, which uses elevated temperature ageing to predict long-term packaging performance and establish expiry dates for pharmaceutical and medical device packaging without waiting for full real-time ageing data.

BINDER heating chambers offer working temperature ranges from ambient up to 300°C depending on the model. Forced-convection models FD 56 and ED 56 deliver temperature uniformity within ±0.5°C to ±1°C throughout the working space. This precision meets ASTM E145 oven uniformity requirements and is essential for reproducible accelerated ageing per ASTM F1980 and conditioning per ASTM D618.

DIN 12880 governs safety requirements for electrically heated laboratory drying ovens, covering overtemperature protection, construction, and electrical safety. For calibration, chambers must meet ISO 17025 traceability requirements. All BINDER ovens include independent overtemperature protection and are designed in full compliance with DIN 12880, ensuring safe and reliable operation in packaging, pharmaceutical, and industrial testing laboratories.

×

Highlights